
Teledyne Analytical Instruments
Part II: 1-3
Thermal Conductivity Analyzer
Part II: Analysis Unit
Figure 1-2: Electrical Connector/Control Panel
•
Power In
Power input terminals for electric heater.
Requires 110 or 220 V ac, depending on
position of the Voltage Selector switch. Use
50/60 Hz.
CAUTION: Check the position of the Voltage Selector switch
BEFORE applying power to the Power Input termi-
nals.
•
Voltage Selector Power input selector switch for electric heater.
Adjusts input requirement for 110 or
220 V ac, depending on available source
voltage. Use 50/60 Hz.
•
Fuses
1.6 A, 250 V, T type, European size
5
×
20 mm fuses. Fuse 2 is on the neutral side
of the line. Fuse 1 is on the hot side of the
line.
•
Solenoid Valves Terminals that provide all electrical intercon-
nections from the Control Unit to the gas
control valves.
•
Sensor Signal
Terminals that provide connections from the
T/C sensor to the Control Unit.
Содержание 2010B
Страница 16: ...Teledyne Analytical Instruments 1 8 Part I 1 Introduction Model 2010B ...
Страница 22: ...2 6 Part I 2 Operational Theory Model 2010B Teledyne Analytical Instruments ...
Страница 34: ...Teledyne Analytical Instruments 3 Installation Model 2010B 3 12 Part I ...
Страница 92: ...4 Maintenance Model 2010B 4 8 Part II Teledyne Analytical Instruments ...