4 Operation
Model 2010B
4-10 Part I
Teledyne Analytical Instruments
RUNNING DIAGNOSTIC
Testing Preamp Cell
When the testing is complete, the results are displayed.
Power: OK Analog: OK
Cell: 2 Preamp: 3
The module is functioning properly if it is followed by
OK
. A number
indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting for number-code information. The results
screen alternates for a time with:
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
4.3.6 The Model Screen
Move the < > arrow key to
MORE
and press
Enter
. With
MODEL
blinking, press
Enter
. The screen displays the manufacturer, model, and soft-
ware version information.
4.3.7 Checking Linearity with ALGORITHM
From the
System
Function screen, select
ALGORITHM
, and press
Enter
.
sel rng to set algo:
> Ø1 Ø2 Ø3 <
Use the < > keys to select the range: 01, 02, or 03. Then press
Enter
.
Gas Use: H2 N2
Range: Ø 10%
Press Enter again.
Algorithm setup:
VERIFY SET UP
Select and
Enter
VERIFY
to check whether the linearization has been
accomplished satisfactorily.
Dpt INPUT OUTPUT
Ø Ø.ØØ Ø.ØØ