3 Installation
Model 2010B
3-6 Part I
To provide an indication of the range, the Range ID analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 3-2 gives the
range ID output for each analysis range.
Table 3-2: Analog Range ID Output—Example
Range
Voltage (V)
Current (mA) Application
Range 1
0.25
8
0-1 % H
2
in N2
Range 2
0.50
12
0-10 % H
2
in N2
Range 3
0.75
16
0-1 % H
2
in Air
Range 4 (Cal)
1.00
20
0-1 % H
2
in N2
3.3.4 Alarm Relays
The three alarm-circuit connectors are spring terminals for making
connections to internal alarm relay contacts. Each provides a set of Form C
contacts for each type of alarm. Each has both normally open and normally
closed contact connections. The contact connections are indicated by dia-
grams on the rear panel. They are capable of switching up to 3 amperes at
250 V ac into a resistive load. See Figure 3-5. The connectors are:
Threshold Alarm 1:
• Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2:
• Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm:
Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as fail-safe and latching. Cannot be de-
feated.
Actuates when cell can not balance during zero
calibration.
Actuates when span parameter out off its limited
parameter.
Содержание 2010B
Страница 16: ...Teledyne Analytical Instruments 1 8 Part I 1 Introduction Model 2010B ...
Страница 22: ...2 6 Part I 2 Operational Theory Model 2010B Teledyne Analytical Instruments ...
Страница 34: ...Teledyne Analytical Instruments 3 Installation Model 2010B 3 12 Part I ...
Страница 92: ...4 Maintenance Model 2010B 4 8 Part II Teledyne Analytical Instruments ...