4 Operation
Model 2000A-EU
4-18
Teledyne Analytical Instruments
•
Value of the alarm setpoint, AL–1 ####
•
Out-of-range direction, HI or LO
•
Defeated? Dft:Y/N (Yes/No)
•
Failsafe? Fs:Y/N (Yes/No)
•
Latching? Ltch:Y/N (Yes/No).
•
To define the setpoint, use the < > arrow keys to move the
blinking over to AL–1 ####. Then use the
∆∇
arrow keys to
change the number. Holding down the key speeds up the
incrementing or decrementing.
•
To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the
∆∇
arrow
keys to change the parameter.
•
Once the parameters for alarm 1 have been set, press
Alarms
again, and repeat this procedure for alarm 2 (AL2).
•
To reset a latched alarm, go to
Dft
and then press either
∆
two
times or
∇
two times. (Toggle it to
Y
and then back to
N.
)
–OR –
Go to
Ltch
and then press either
∆
two times or
∇
two times.
(Toggle it to
N
and back to
Y.
)
4.6
The
Range Select Function
The
Range
function allows you to manually select the concentration range
of analysis (
MANUAL
), or to select automatic range switching (
AUTO
).
In the
MANUAL
screen, you are further allowed to define the high and low
(concentration) limits of each Range, and select a single, fixed range to run.
CAUTION: If this is a linearized application, the new range must
be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.
In the
AUTO
screen, you are further allowed to select which gas applica-
tion (PREVIOUSLY defined in
System
function) to run.