Theory Of Operation
Model 100E Instruction Manual
190
045150102 Rev XB1
10.6. Communications Interface
The analyzer has several ways to communicate the outside world, see Figure 10-19.
Users can input data and receive information directly through the front panel keypad and
display. Direct, two-way communication with the CPU is also available by way of the
analyzer’s RS232 & RS485 I/O ports (see Section 6.9 and 6.11). Alternatively, an
Ethernet communication option can be substituted for one of the Comm ports.
The analyzer can also send status information and data via the eight digital status output
lines (see Section 6.8.1) and the three analog outputs (see Section 6.5) located on the
rear panel as well as receive commands by way of the six digital control inputs also
located on the rear pane (see Section 6.8.2).
Figure 10-19: Interface Block Diagram
Содержание 100E
Страница 2: ...Model 100E Instruction Manual ...
Страница 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Страница 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Страница 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Страница 25: ...Model 100E Instruction Manual Specifications Approvals and Warranty 045150102 Rev XB1 23 User Notes ...
Страница 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Страница 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Страница 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Страница 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...
Страница 252: ...APPENDIX A 6 Terminal Command Designators Revision B 3 Model 100E Instruction Manual 250 045150110 Rev B 3 ...
Страница 258: ...Appendix D Electronic Schematics M100E Model 100E Instruction Manual 256 045150140 Rev XB1 User Notes ...