EPA Protocol Calibration
Model 100E Instruction Manual
152
045150102 Rev XB1
Follow the procedures described in section 7.2 for calibrating the zero points.
For each mid point:
SPAN CAL M
RANGE = 500.0 PPB
SO
2
X.XXX
< TST TST >
ZERO
SPAN
CONC
EXIT
Wait until
STABIL
falls
below
0.5 ppb.
This may take
several minutes.
SAMPLE
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST > CAL
CALZ
CALS
SETUP
Record the
SO
2
\reading as
displayed on the
instrument’s front
panel
Press
EXIT
to
Return to the
Main
SAMPLE
Display
ACTION:
Allow Calibration Gas diluted to
proper concentration for
Midpoint N+1 to enter the sample
port
ACTION:
Allow calibration gas diluted to proper concentration for
Midpoint N to enter the sample port
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST
TST
>
CAL
SETUP
SAMPLE
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
CAL
SETUP
Set the Display to show the
STABIL
test function.
This function calculates the
stability of the NO/NO
x
measurement
8.6. Special Calibration Requirements for Dual Range or
Auto Range
If Dual Range or Auto Range is selected, then it should be calibrated for both Range1 and
Range2 separately.
For zero and span point calibration, follow the procedure described in Section 7.2. Repeat
the procedure for both the
HIGH
and
LOW
Ranges
8.7. References
1. Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 50,
Appendix A, Section 10.3.
Содержание 100E
Страница 2: ...Model 100E Instruction Manual ...
Страница 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Страница 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Страница 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Страница 25: ...Model 100E Instruction Manual Specifications Approvals and Warranty 045150102 Rev XB1 23 User Notes ...
Страница 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Страница 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Страница 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Страница 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...
Страница 252: ...APPENDIX A 6 Terminal Command Designators Revision B 3 Model 100E Instruction Manual 250 045150110 Rev B 3 ...
Страница 258: ...Appendix D Electronic Schematics M100E Model 100E Instruction Manual 256 045150140 Rev XB1 User Notes ...