Operating Instructions
Model 100E Instruction Manual
104
045150102 Rev XB1
Table 6-20: iDAS Data Channel Properties
Property Description
Default
Setting
Range
NAME
The name of the data channel.
“NONE”
Up to 6 letters and
digits (more with
APICOM, but only the
first six are displayed
on the front panel).
TRIGGERING
EVENT
The event that triggers the data channel to
measure and store its data parameters. See
APPENDIX A-5 for a list of available triggering
events.
ATIMER
Any allowed event.
NUMBER AND
LIST OF
PARAMETERS
A User-configurable list of data types to be
recorded in any given channel. See APPENDIX
A-5 for a list of available parameters
1 -
PMTDET
Any available
concentration,
temperature,
pneumatic or
diagnostic parameter.
REPORT PERIOD The amount of time between each channel data
point.
000:01:00 000:00:01
to
366:23:59
(Days:Hours:Minutes)
NUMBER OF
RECORDS
The number of reports that will be stored in the
data file. Once the specified limit has been
exceeded, the oldest data are over-written to
make space for new data.
100
1 to 1 million, limited
by available storage
space.
RS-232 REPORT Enables the analyzer to automatically report
channel values to the RS-232 ports.
OFF
OFF or ON
CHANNEL
ENABLED
Enables or disables the channel. Provides a
convenient means to temporarily disable a data
channel.
ON
OFF or ON
CAL HOLD OFF
Disables sampling of data parameters while
instrument is in calibration mode.
Note that - when enabled here - there is also a
length of the DAS HOLD OFF after calibration
mode, which is set in the VARS menu (Section
6.10.2.11.)
OFF
OFF or ON
6.10.1.2. iDAS Parameters
Data parameters are types of data that may be measured and stored by the iDAS. For
each T-API analyzer model, the list of available data parameters is different, fully defined
and not customizable. Appendix A-5 lists firmware specific data parameters for the
M100E. The most common parameters are concentrations of measured gases (SO
2
, SO
2
and NO
2
), temperatures of heated zones (converter, sample chamber, box
temperature…), pressures and flows of the pneumatic subsystem and other diagnostic
measurements as well as calibration data (slope and offset) for each gas.
Most data parameters have associated measurement units, such as mV, ppb, cm³/min,
etc., although some parameters have no units. The only units that can be changed are
those of the concentration readings according to the
SETUP-RANGE
settings. Note that
the iDAS does not keep track of the unit of each concentration value and iDAS data files
may contain concentrations in multiple units if the unit was changed during data
acquisition.
Содержание 100E
Страница 2: ...Model 100E Instruction Manual ...
Страница 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Страница 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Страница 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Страница 25: ...Model 100E Instruction Manual Specifications Approvals and Warranty 045150102 Rev XB1 23 User Notes ...
Страница 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Страница 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Страница 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Страница 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...
Страница 252: ...APPENDIX A 6 Terminal Command Designators Revision B 3 Model 100E Instruction Manual 250 045150110 Rev B 3 ...
Страница 258: ...Appendix D Electronic Schematics M100E Model 100E Instruction Manual 256 045150140 Rev XB1 User Notes ...