background image

PP019/PP020 Passive Probe 

EAR99 technology subject to restrictions on copyrights page. 

Do not operate with suspected failures. Before each use, inspect the probe 

and accessories for any potential damage such as tears or other defects in 

the probe body, cable jacket, accessories, etc. If any part is damaged, cease 

operation immediately and sequester the probe from inadvertent use. 

Cleaning 

 

Clean only the exterior of the probe, cables, and accessories. Use a 

soft cotton cloth moistened with a mild detergent and water solution. 

Do not allow any portion of the probe to be submerged at any time. 

 

Dry the probe and accessory thoroughly before attempting to make 

any voltage measurement. 

 

Do not subject the probe to solvents or solvent fumes as these can 

cause deterioration of the probe body, cables, and accessories. 

Service 

Refer all repairs to qualified service personnel. Contact Teledyne LeCroy at 

the following address to arrange to return the probe for service: 

Teledyne LeCroy, Inc. 

700 Chestnut Ridge Road 

Chestnut Ridge, NY, 10977-6499 

Ph: 800-553-2769 or 845-425-2000 

Fax: 845-578-5985 

[email protected] 

For the most complete and up-to-date list of sales and service centers by 

country, visit teledynelecroy.com/support/contact. 

Содержание PP019

Страница 1: ...User Manual PP019 and PP020 Passive Probes...

Страница 2: ...e LeCroy documentation materials is strictly prohibited Information contained herein is classified as EAR99 under the U S Export Administration Regulations Export reexport or diversion contrary to U S...

Страница 3: ...PP020 are passive high impedance voltage oscilloscope probes designed and calibrated for use with general purpose oscilloscopes having an input impedance of 1 M shunted by 16 pF However they may be co...

Страница 4: ...t exceed the voltage or category ratings of the probe See voltage derating curve on p 4 PollutionDegree PollutionDegree2 OperatingAltitude Upto3000meters Max OperatingTemp 0to50 C Humidity 5 to 95 RH...

Страница 5: ...User Manual EAR99 technology subject to restrictions on copyrights page 3 PP019 Input Impedance Profile PP020 Input Impedance Profile...

Страница 6: ...Official Journal of the European Communities IEC EN 61010 031 2015 Safety requirements for electrical equipment for measurement control and laboratory use Part 031 Safety requirements for handheld pro...

Страница 7: ...t Disconnect the probe input and reference lead from the test circuit before disconnecting from the oscilloscope Do not connect disconnect probes while connected to a voltage source Do not overload Do...

Страница 8: ...ution Do not allow any portion of the probe to be submerged at any time Dry the probe and accessory thoroughly before attempting to make any voltage measurement Do not subject the probe to solvents or...

Страница 9: ...d be adjusted whenever the probe is connected to or transferred between oscilloscopes Low Frequency PP019 and PP020 1 Apply a 1 kHz square wave to the probe or connect to the oscilloscope s calibrator...

Страница 10: ...nt However if adjustment is required 1 Apply a 1 MHz square wave to the probe 0 7 ns rise time 2 Remove the plastic caps from the two HF trimmers on the PP020 BNC Compensation Box HF trimmers on PP020...

Страница 11: ...ion Quantity 1 Sprung Hook 5 mm Black 1 2 GND Lead w Alligator Clip 1 3 BNC Adapter 5 mm 1 4 Replacement Tip Black 1 5 Deluxe Trimmer Tool 1 6 IC Tip Insulator 5 mm Black 1 7 Tip Insulator 5 mm Black...

Страница 12: ...rranty Teledyne LeCroy at its option will either repair or replace any assembly returned within the warranty period to the Customer Service Department of an authorized service center However this will...

Страница 13: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Teledyne LeCroy PP019 1 PP020 1...

Отзывы: