PP019/PP020 Passive Probe
2
EAR99 technology subject to restrictions on copyrights page.
Specifications
Model
Attn.
Input
Impedance
B/W
Rise
Time
Length
Compensation
Range
R (MΩ)
C (pF)
(MHz)
(ns)
(m)
(pF)
PP019
10x
10
12
250
1.4
1.2
10 - 30
PP020
10x
10
11
500
0.7
1.2
8 - 20
Attenuation Ratio .................. 10:1 ±1%
Max. Input Voltage ....... Circuits not directly connected to Mains: 500 Vrms,
......................................... 1500 V transient overvoltage
......................................... Measurement Category II*: 400 Vrms
To reduce risk of fire or electric shock, do not exceed the voltage or
category ratings of the probe. See voltage derating curve on p.4.
Pollution Degree ..................... Pollution Degree 2*
Operating Altitude ................. Up to 3000 meters
Max. Operating Temp ......... 0 to 50 °C
Humidity ....................................
5% to 95% RH (10 to 30 °C)
5% to 75% RH (30 to 40 °C)
5% to 50% RH (above 40 °C)
RH not controlled below 10 °C
* As defined in IEC/EN 61010-031. See definitions below.
Definitions:
Measurement Category II (CAT II) refers to measurements performed on circuits directly connected to
utilization points (socket outlets and similar) of the low-voltage mains installation.
Pollution Degree 2 refers to an operation environment where normally only dry non-conductive
pollution occurs. Temporary conductivity caused by condensation must be expected.