Operator’s Manual
61
Appendix 1: OMA Calibration
The OMA system is calibrated at various stages to enable highly accurate signal measurements.
Coherent Receiver Factory Calibrations
•
Skew between output channels of the coherent receiver including external cables
•
Frequency Response of the coherent receiver
•
Temperature dependence of the coherent receiver
Full System Factory Calibrations
•
X-pol and Y-pol channel balance of the complete system, including the oscilloscope
Manual Calibrations
•
LabMaster ChannelSync (deskew) calibration
Dynamic Calibrations
•
LabMaster temperature-dependent calibration (
±
5
°
C factory calibration temperature)
•
OMA software calibrations prior to each signal measurement. These dynamic calibrations include:
o
Voltage scale optimization
o
I & Q channel imbalance correction
o
Offset removal between all channels
o
Channel deskew (fine-tuning in sub-ps scale, to line up sampling points)