083730510A DCN8172
Teledyne API T500U CAPS NO
2
Analyzer with NumaView™ Software
33
BASIC CONNECTIONS FROM CALIBRATOR
VENT
here if input
is pressurized
Source of
SAMPLE GAS
Removed during
calibration
Chassis
SAMPLE
EXHAUST
MODEL T700
Gas Dilution
Calibrator
VENT
MODEL 701
Zero Gas
Generator
Enclosure Wall
Calibrated NO
2
at
HIGH
Span
Concentration
Figure 2-13. Gas Line Connections from Calibrator – Basic Configuration
For the analyzer’s basic configuration, in addition to tubing, pressure, venting, and exhaust
requirements set out in Section 2.3.2.1, attach the following pneumatic lines:
S
AMPLE
G
AS
S
OURCE
Connect a sample gas line to the SAMPLE inlet.
C
ALIBRATION
G
AS
S
OURCES
CAL GAS & ZERO AIR SOURCES: The source of calibration gas is attached to the
SAMPLE inlet, but only when a calibration operation is actually being performed.
V
ENTING
Vent the output of the calibrator if calibrator not already vented.
E
XHAUST
O
UTLET
Attach an exhaust line to the EXHAUST outlet fitting and vent outside the shelter.
Содержание T500U