Teledyne API T500U CAPS NO
2
Analyzer with NumaView™ Software
083730510A DCN8172
CRITICAL TUBING, PRESSURE, VENTING AND EXHAUST
REQUIREMENTS
The requirements presented in this section apply to all pneumatic connection instructions.
Tubing
:
•
PTFE material
•
Outer diameter (OD) minimum ¼”
•
Min/max length 2 meters to 10 meters
Pressure
:
•
All Sample gas pressure must be at ambient atmospheric pressure, no greater than
1.0 psig.
Venting
(to prevent back diffusion and pressure effects):
•
Run tubing outside the enclosure or at least away from immediate area surrounding
the instrument.
Exhaust Outlet
:
•
Run tubing outside the enclosure.
Calibration Gas Sources
:
•
The source of calibration gas or zero air is also attached to the
SAMPLE
inlet, but
only when a calibration operation is actually being performed.