1. INTRODUCTION
The T108 consists of two major assemblies: a modified T100 SO
2
analyzer and a 501TS thermal
converter. This manual addendum describes the specifics of the T108 Analyzer that differ from the T100
Analyzer; all other user information is contained in the T100 manual, PN 06807.
1.1. SPECIFICATIONS
The specifications and the warranty for the SO
2
analyzer are contained in the T100 manual. However, the
AC power specifications for the T108 differ as follows:
T108 AC Power:
100V – 120V, 60Hz (205W); 220V – 240V, 50Hz (215W)
The specifications for the 501TS Converter are presented in Table 1-1
Table 1-1. 501TS Converter Specifications
Specification
Value
Maximum Flow Rate
1000 cc/min
Nominal Flow Rate (CO
2
)
625 cc/min
Nominal Flow Rate (Air/N
2
)
450 cc/min
Maximum TS Concentration for
specified conversion efficiency
20 ppmv
Minimum Conversion Efficiency
(In CO
2
matrix)
H
2
S
COS, CS
2
98%
90%
Least Discernible Level (LDL)
See T100 Manual
Operating Converter Temperature
1000
o
C
Maximum Converter Temperature
1050
o
C
Power
100-120/220-240 VAC
50/60 Hz, (440 W)
Weight
24 lbs (11kg)
Dimensions
7in x 17in x 22in
(178mm x 432mm x 559mm)
1.2. THE T108 TOTAL-SULFIDES-IN-CO
2
ANALYZER
The Teledyne API Model T108 Total Sulfides in CO
2
Analyzer, is designed to measure mixed sulfur
impurities, collectively referred to as Total Sulfides (TS), in carbon dioxide (CO
2
) gas. Since there is no
SO
2
scrubber in the system, the instrument reading is the sum of the reduced sulfur compounds and SO
2
.
The T108 consists of a modified T100 UV Fluorescence SO
2
Analyzer, with special software, and a
501TS high temperature quartz thermal converter.
The 501TS primarily consists of a heated, temperature-controlled quartz tube. Sulfur compounds are
heated to approxima
tely 1000 ˚C as they pass through the quartz tube and are converted to SO
2
in the
following manner:
TS + O
2
--> SO
2
Содержание T108 Series
Страница 2: ......
Страница 4: ...Teledyne API Model T108 Addendum to T100 Manual ii 07268C DCN8258 This page intentionally left blank...
Страница 6: ...Teledyne API Model T108 Addendum to T100 Manual iv 07268C DCN8258 This page intentionally left blank...
Страница 26: ...Appendix A Model 501 Interconnects 07268C DCN8258...