2-5
P
P
P
P
Per
er
er
er
ercent Oxyg
cent Oxyg
cent Oxyg
cent Oxyg
cent Oxygen Anal
en Anal
en Anal
en Anal
en Analyz
yz
yz
yz
yzer
er
er
er
er
Operational Theory 2
Teledyne Analytical Instruments
Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel Cell
2.2.6 Micro-Fuel Cell “Class”
TBE manufactures Micro-Fuel Cells with a variety of characteristics to
give the best possible performance for any given sample conditions. A few
typical Micro-Fuel Cells are listed below with their typical use and electrical
specifications.
2.2.6.1 Class A-3 Cell
The class A-3 cell is for use in applications where it is exposed continu-
ously to carbon dioxide concentrations between 1 % and 100 % in the
sample gas.
Nominal output in air is 0.20 mA, and 90 % response time is 45 s.
Expected life in flue gas is 8 months.
2.2.6.2 Class A-5 Cell
The class A-5 cell is for use in applications where it is exposed intermit-
tently to carbon dioxide concentrations up to 100 % in the sample gas.