Percent Oxygen Analyzer
Operation 4
4-1
Teledyne Analytical Instruments
Operation
4.1
Introduction
Once the analyzer has been installed, it can be configured for your
application. To do this you will:
•
Set system parameters:
•
Establish a security password, if desired, requiring Operator
to log in.
•
Establish and start an automatic calibration cycle, if desired.
•
Calibrate the instrument.
•
Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.
•
Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).
Before you configure your 3000PA these default values are in effect:
Ranges: LO = 1 %, MED = 5 %, HI = 10 %
Auto Ranging: ON
Alarm Relays: Defeated, 10 %, HI, Not failsafe, Not latching
Zero: Auto, every 0 days at 0 hours
Span: Auto, at 20.9 %, every 0 days at 0 hours
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press
Enter
for access to all functions of the analyzer.
Содержание 3000PA-EU
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 56: ...Percent Oxygen Analyzer Operation 4 4 19 Teledyne Analytical Instruments...
Страница 57: ...4 Operation 4 Operation 4 Operation 4 Operation 4 Operation Model 3000PA 4 20 Teledyne Analytical Instruments...
Страница 77: ...A 10 Appendix Appendix Appendix Appendix Appendix Model 3000PA Teledyne Analytical Instruments...
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......