Control Solution Troubleshooting
If your control solution testing is out of range (too high or too low),
it may be caused by the following:
n
Wrong brand of control solution.
n
Control solution not at room temperature.
n
Expired or contaminated control solution.
n
Damaged test strip.
n
Touched the tip of solution bottle to test strip.
n
Meter malfunction.
n
Wrong ambient temperature (should be between 50˚F and 104˚F).
n
Check the expiration date and open bottle date on the control solution
and on the test strips. Repeat testing using a new test strip. If the result is
still out of range, use a new bottle of control solution and retest.
If the problem persists, turn your meter off and then call TELCARE™
at (888) 666-4530. Do not perform any blood glucose tests until you
resolve this issue.