TELCARE
™
Blood Glucose Monitoring System
3. Error Message:
Not Enough Blood
What it means:
There is not enough blood or too much blood or control solution
on the test strip to give a reliable reading.
Action to take:
Remove the test strip and repeat the test with a new test strip.
4. Error Message:
Temp. Too Low / Temp. Too High
What it means:
The meter is too hot or too cold.
Action to take:
Move the meter into an area that is between 50˚F and 104˚F.
IMPORTANT:
After receiving this error, allow the meter to be at room temperature for at
least 30 minutes before testing blood or control solution.
38
If your meter displays a “Low Battery”
icon, charge the battery.
Other Messages
High / Low Displayed
If you see either High or Low displayed on
your device, your blood sugar level may be
out of range (either in excess of 600 mg/dL
[High] or lower than 20 mg/dL [Low]).