![Tektronix WVR6100 Series Скачать руководство пользователя страница 86](http://html1.mh-extra.com/html/tektronix/wvr6100-series/wvr6100-series_instructions-manual_1078279086.webp)
Incoming Inspection
2- 32
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
1.
Connect an Ethernet cable from the rear of the unit to a computer with a Web
browser.
2.
Press FACTORY to restore the factory preset.
3.
Press CONFIG to display the Configuration menu.
4.
Select Network Settings, and press the right-arrow key to change the menu
focus to the submenu.
5.
Verify the IP Config Mode is set to Manual.
6.
Select IP Address, and press the right-arrow key to change the menu focus to
the submenu.
7.
Use the left/right arrow keys to move the highlight to each box in the
address. Use the up/down arrow keys to change the number in each box as
required. Configure the IP Config Mode and IP Address to be compatible
with the computer.
8.
To accept your changes, move the highlight to the Accept box and press
SEL.
9.
From the computer, enter the IP address of the waveform rasterizer into the
Web browser address line (for example, http://192.182.256.23).
10.
You should see a Web page that lists several options for accessing the
waveform rasterizer from the PC. If this happens, then the Ethernet function
is working.
11.
Record Pass or Fail for Ethernet in the test record.
Ethernet Test
Содержание WVR6100 Series
Страница 4: ......
Страница 12: ...Preface viii WVR6100 WVR7000 and WVR7100 Specifications and Performance Verification ...
Страница 54: ...Specifications 1 42 WVR6100 WVR7000 and WVR7100 Specifications and Performance Verification ...
Страница 56: ...Performance Verification 2 2 WVR6100 WVR7000 and WVR7100 Specifications and Performance Verification ...
Страница 142: ...Audio Performance Verification 2 88 WVR6100 WVR7000 and WVR7100 Specifications and Performance Verification ...