This test checks the SD jitter waveform accuracy.
1.
Set tile 2 for Jitter display as follows:
a.
Press the DISPLAY SELECT 2 button.
b.
Press and hold EYE (or MEAS if no EYE button), and then select
Display Type > Jitter.
c.
Press MAG and FIELD for a sweep rate of 150 us/div.
2.
Adjust the DVG1 SDI module parameters as follows:
a.
Set the jitter frequency to 9950 Hz.
b.
Slowly increase the jitter amplitude while watching the Eye display. As
the jitter amplitude increases, the transitions start to get wider and fill in
the space between the transitions. As the jitter amplitude approaches one
UI, the space between transitions disappears (the eye will close). At one
UI of jitter the eye diagram will have “closed”, and a darkening, or
shadow, on an eye trace will appear out of the smeared background.
c.
Continue increasing the jitter until another shadow of an eye appears.
You now have 2 UIs of jitter.
d.
Continue increasing jitter to 3 UIs. Do not adjust for 3 UIs on the
generator; adjust instead for the sharpest narrow eye shadow on the
rasterizer display, which indicates that the applied jitter is precisely 3UI.
This may not be 3 UI on the generator because you are setting it for 3 UI
added jitter instead of 3 UI total jitter.
3.
The jitter display should show a nearly stationary sine--wave 3 UI in height.
Check the sine portion of the jitter trace, ignoring the small random noise
component, with the graticule marks and the cursors to see that it is 3 UI
ᐔ
10% (between 2.7 UI and 3.3 UI).
4.
Record Pass or Fail in the test record for SD Jitter Waveform Accuracy.
This test checks that the SD recovered clock signal has no fidelity issues (such as
stuck bits), after it has been filtered into a sine-wave and digitized.
1.
Press FACTORY.
2.
Connect a DVG1 output to the SDI A input. Set the DVG1 to generate a
525--270 100% Color Bars signal.
3.
Terminate the other SDI A connector with a 75
Ω
SDI terminator.
4.
Press CONFIG, and then select Utilities > Calibration. Press SEL to start
calibration.
SD Jitter Waveform
Accuracy
(WVR6100 Option PHY,
WVR7100 Options SD and
PHY)
SD Recovered Sine-wave
Fidelity
(WVR6100 & Option EYE
or PHY, WVR7100 Option
SD and EYE or PHY)
Содержание WVR6100 Series
Страница 4: ......
Страница 12: ...Preface viii WVR6100 WVR7000 and WVR7100 Specifications and Performance Verification ...
Страница 54: ...Specifications 1 42 WVR6100 WVR7000 and WVR7100 Specifications and Performance Verification ...
Страница 56: ...Performance Verification 2 2 WVR6100 WVR7000 and WVR7100 Specifications and Performance Verification ...
Страница 142: ...Audio Performance Verification 2 88 WVR6100 WVR7000 and WVR7100 Specifications and Performance Verification ...