Audio Performance Verification
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
2- 83
11.
Set the analog audio analyzer to measure RMS level in dBu, and for an input
impedance of 200 k
Ω
. This is done on the R&S UPL06 as follows:
a.
Press ANLR to display and configure the ANALYZER panel.
b.
Use the up/down arrow keys, keypad, and SELECT button to set the
parameters in the ANALYZER panel as follows:
INSTRUMENT
ANLG 22 kHz
Min Freq
10 Hz
Ref Imped
100000
Ω
Channel
1
Ch1 Coupl
AC
Ch1 Input
BAL
Ch1 Imped
200 k
Ω
Ch1 Common
FLOAT
Ch1 Range
AUTO
START COND
AUTO
Delay
0.0000 s
INPUT DISP
OFF
FUNCTION
RMS & S/N
S/N Sequ
OFF
Meas Time
AUTO
Unit Ch1
DBu
Reference
VALUE: 1.0000 V
Sweep Mode
NORMAL
Notch (Gain)
OFF
Filter
OFF
Filter
OFF
Filter
OFF
12.
Connect the R&S UPL06 Opt B22, B29 digital generator unbalanced
(UNBAL) output to the AES A 1--2 In BNC using a 75
Ω
cable.