Incoming Inspection
Dual Link Format
Veri
fi
cation Test (Option
DL)
1.
Set up a dual link signal generator with an HDLG7 module. Connect the Link
A output on the module to the SDI A input on the instrument, and the Link B
output on the module to the SDI B input on the instrument.
2.
Connect the loopthrough terminators on the instrument.
3.
Perform the following steps to set the HDLG7 module format to 1080 59.94i:
a.
On the signal generator, press the
FORMAT
button until 1080 59.94i
appears, and then press the
ENTER
button.
b.
Press the
CANCEL
button.
c.
Press the up/down arrows until Sample Structure/Depth appears.
d.
Press the left/right arrows until YCbCr+A 4:4:4 10 bits appears.
e.
Press the
ENTER
button.
f.
Press the
COLOR BAR
button until 75% color bars appears.
4.
Press the
FACTORY
button on the instrument under test to restore factory
settings.
5.
Press
DISPLAY SELECT 2
to select tile 2.
6.
Press and hold the
STATUS
button.
7.
Select
Display Type > Video Session
.
8.
Press
DISPLAY SELECT 4
to select tile 4.
9.
Press and hold the
VECT
button to display the Vector map and the Vector
menu.
10.
Select
Display Type > Lightning
.
11.
Press the
VECTOR
button to close the menu.
70
Waveform Rasterizers Speci
fi
cations and Performance Veri
fi
cation