Speci
fi
cations
Table 20: Environmental characteristics (cont.)
Characteristic
Performance requirement
Reference information
Power Spectral Density
0.00015 g2/Hz from 5 to 350 Hz.
–3 dB/octave from 350 to 500 Hz.
0.000105 g2/Hz at 500 Hz.
Overall level of 0.27 g
RMS
, 10 minutes per
axis.
Non-operating
Tektronix Class 5 Random Vibration
Non-operating Test: For Laboratory /
Benchtop Products.
Power Spectral Density
0.0175 g2/Hz from 5 to 100 Hz.
–3 dB/octave from 100 to 200 Hz.
0.00875 g2/Hz from 200 to 350 Hz.
–3 dB/octave from 350 to 500 Hz.
0.006132 g2/Hz at 500 Hz.
Overall level of 2.28 g
RMS
, 10 minutes per
axis.
Mechanical Shock, typical
Non-operating
30 g 1/2 sine, 11 ms duration, 3 shocks per
surface (18 total).
Meets Mil-PRF-28800F Class 3.
Transportation, typical
TEK Std 062-9537-00.
Meets International Safe Transit Association
Test Procedure 1A, April 1996, Category
II for Vibration, Impact (24 inch drop), and
Compression.
High De
fi
nition (HD) Characteristics
The characteristics in this section only apply to instruments with Option HD.
Table 21: HD waveform vertical characteristics, SDI inputs A and B
Characteristic
Performance requirement
Reference information
Frequency Response
Luminance channel (Y) to
30 MHz (to 60 MHz for 1080P
60/59/50 dual link formats)
± 0.5%
RGB Monochrome identical to Y channel.
Color difference channels (P
B
& P
R
) to 15 MHz (to 30 MHz
for 1080P 60/59/50 dual link
formats)
± 0.5%
Transient Response
Pulse-to-bar ratio 0.99:1 to 1.01:1 on
appropriate Sine-Squared or Blackman 2T
pulse.
12
Waveform Rasterizers Speci
fi
cations and Performance Veri
fi
cation