Performance Verification
WFM 601 - Service Manual
4–21
i.
Turn on the CH B display.
j.
Adjust the Reference Level for a screen display.
k. Turn off spectrum analyzer CH A and CH D display.
l.
Note the reference level of the straight trace.
m. Select B Save A.
n. Remove the bnc female-to-female connector coupling the 50
-to-75
minimum loss attenuators together.
o. Connect the two cables, with the 50
-to-75
minimum loss attenuators
to the WFM
601 SER A INPUT.
p. CHECK – that the spectrum analyzer trace is within 1 major division (1
dB) of the reference level established in step l. (Channel selected or
not.)
q. Move the cables and the 50
-to-75
minimum loss attenuators to the
WFM
601 SER B INPUT.
r.
CHECK – that the spectrum analyzer trace is within 1 major division (1
dB) of the reference level established in step l. (Channel selected or
not.)
s.
Reverse cable to the WFM
601 SER B INPUT.
t.
CHECK – that the spectrum analyzer trace is within 1 major division (1
dB) of the reference level established in step l. (Channel selected or
not.)
This completes the Performance Check procedure.
Содержание WFM601
Страница 4: ...The WFM601 Serial Digital Component Monitor...
Страница 10: ...Contents iv WFM 601 Service Manual Revised Nov 1994...
Страница 14: ...Contents viii WFM 601 Service Manual Revised Nov 1994...
Страница 19: ......
Страница 32: ......
Страница 50: ...Installation 2 18 WFM 601 Service Manual...
Страница 57: ......
Страница 87: ...Theory of Operation 3 30 WFM 601 Service Manual...
Страница 88: ......
Страница 110: ...Performance Verification 4 22 WFM 601 Service Manual...
Страница 111: ......
Страница 132: ......
Страница 168: ...Maintenance 6 36 WFM 601 Service Manual...
Страница 169: ......
Страница 175: ...Options 7 6 WFM 601 Service Manual...
Страница 176: ......
Страница 177: ......
Страница 226: ......
Страница 233: ...WFM 601 SERIAL COMPONENT MONITOR BNC INPUT BOARDS 1...
Страница 237: ...ECL TO TTL CONVERTER LINE BUFFER RAM WFM 601 SERIAL COMPONENT MONITOR COPROCESSOR 3...
Страница 241: ...2 1 10 4256403 Y DAC DIGITAL HALF BAND FILTER Y DELAY WFM 601 SERIAL COMPONENT MONITOR Y DELAY HALF BAND FILTER DAC 5...
Страница 245: ...Y FILTER Pb FILTER Pr FILTER WFM 601 SERIAL COMPONENT MONITOR Y Pb Pr RECONSTRUCTION FILTERS 7...
Страница 247: ...WFM 601 SERIAL COMPONENT MONITOR TRANSCODERS PIX MONITOR OUTPUTS 8...
Страница 249: ...0 0 0 0 WFM 601 SERIAL COMPONENT MONITOR LIGHTNING VECTOR BOWTIE SWITCHING 9...
Страница 251: ...WFM 601 SERIAL COMPONENT MONITOR CONTROL DACULATOR 10...
Страница 257: ...HORIZONTAL DEFLECTION SWEEP GENERATORS HORIZONTAL SIGNAL MUX HORIZONTAL AMPLIFIER WFM 601 SERIAL COMPONENT MONITOR...
Страница 259: ...ROM MICROPROCESSOR NOVRAM RAM FLASH EPROM BUFFERS RESET ABORT WFM 601 SERIAL COMPONENT MONITOR...
Страница 261: ...OUTPUTS SYNCHRONOUS LINE RATE CONTROLLER SYNC SEPARATOR WFM 601 SERIAL COMPONENT MONITOR DYNAMIC CONTROL 15...
Страница 265: ......
Страница 271: ...WFM 601 SERIAL COMPONENT MONITOR FRONT PANEL 20...
Страница 282: ...WFM 601 FIG 1 EXPLODED VIEW A1 A2 A3 A4 A4A1 A5 A6 A7 A8...
Страница 290: ...Glossary Glossary 8 WFM601 Serial Component Monitor...
Страница 304: ...A8 DAC Board 671 2676 04 M81119 Page 4 of 4...