cation procedures

Veri

fi

cation procedure

Complete the following steps to complete this procedure. Record the results
on the Calibration Data Sheet.

1.

Set the DTG sequencer to RUN and the outputs ON.

2.

Press the RUN button on the TLA and wait a few seconds to verify that it
does not trigger.

3.

Starting from 0.000 ns, increase the delay of the DTG clock channel in 100 ps
steps until triggering begins to occur. When the TLA begins to trigger,
decrease the delay in 10 ps steps to

fi

nd the trigger threshold to within 10 ps.

Record this delay amount.

Note that the logic analyzer might trigger because of a glitch when you make
a delay change. If the data in the waveform window is correct (all data
transitioning at the same time and at the correct frequency), then ignore this
"false trigger" and start the logic analyzer again.

As an alternative, you may want to run the logic analyzer in continuous loop
mode if the DTG causes a false trigger on the logic analyzer each time you
change the delay. Then observe if the data is correct in the waveform window
and ignore any false triggers. Continue increasing the clock delay until the
waveform window displays data that was not acquired correctly. Record this
delay.

4.

Add 300 ps to the delay value that you recorded in step 3 and increase the
DTG clock delay to match this cumulative value. (For example, if you
measured 600 ps, increase the delay to 900 ps.)

5.

Press Run and wait a few seconds to verify that it does not trigger. This
veri

fi

es that the setup and hold window is less than 300 ps, which is the

guaranteed speci

fi

cation for a single channel.

To measure the actual setup and hold window size for your application, slowly
decrease the clock delay in steps (waiting a few seconds between steps to
verify that it does not trigger), until the logic analyzer triggers. Record this
second value. The difference between this second value and the value that you
measured in step 3 is the measured setup and hold window size.

24

TLA6400 Series Product Speci

fi

cations & Performance Veri

fi

cation

Содержание TLA6400 Series

Страница 1: ...ent supports TLA Application Software V6 0 and above Warning These servicing instructions are for use by qualified personnel only To avoid personal injury do not perform any servicing unless you are q...

Страница 2: ...is publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc MagniVu and iView a...

Страница 3: ...tion procedures 13 Summary verification 13 Test equipment 14 Functional verification 14 Certification 16 Performance verification procedures 16 Threshold accuracy test 17 Setup and hold 21 Test fixtur...

Страница 4: ...Table 2 External signal interface 2 Table 3 PC controller 4 Table 4 External PC controller interfaces 4 Table 5 Display system 5 Table 6 Front panel interface 5 Table 7 AC power source 5 Table 8 Prob...

Страница 5: ...ected to earth ground Before making connections to the input or output terminals of the product ensure that the product is properly grounded Observe all terminal ratings To avoid fire or shock hazard...

Страница 6: ...practices that could result in injury or loss of life CAUTION Caution statements identify conditions or practices that could result in damage to this product or other property Symbols and terms on the...

Страница 7: ...ndering first aid and resuscitation is present Disconnect power To avoid electric shock switch off the instrument power then disconnect the power cord from the mains power Use care when servicing with...

Страница 8: ...Service safety summary vi TLA6400 Series Product Specifications Performance Verification...

Страница 9: ...fer to the Tektronix Web site www tektronix com manuals Related documentation Item Purpose TLA Quick Start User Manuals High level operational overview Online Help In depth operation and UI help Insta...

Страница 10: ...Preface viii TLA6400 Series Product Specifications Performance Verification...

Страница 11: ...s The instrument must be in an environment with temperature altitude humidity and vibration within the operating limits described in these specifications The instrument must have had a warm up period...

Страница 12: ...emperature 29 C 84 2 F derates relative humidity to approximately 22 at 50 C 122 F and 10 at 60 C 140 F Operating To 3000 m 9843 ft derated 1 C 34 F per 300 m 984 ft above 1500 m 4921 ft altitude Nono...

Страница 13: ...put voltage Vth 900 mV Input mode Active true low level sensitive Signal 1 2 Signal 3 4 Input bandwidth 1 50 MHz square wave minimum 10 MHz square wave minimum External Signal Input Active period Acce...

Страница 14: ...scription CPU Intel Core i3 2120 3 3 GHz 3 M cache DRAM 2 x 2 GB DIMM 4 GB total DDR3 1066 MHz PC3 8500 Hard disk drive 500 GB or greater SATAII 3 5 inch 7200 RPM Optical drive Optical CD RW DVD R RW...

Страница 15: ...lays only Horizontal position Scrolling and positioning dependent on display type Special function knobs Horizontal scale Scales waveform displays only Table 7 AC power source Characteristic Descripti...

Страница 16: ...x 5 x P5910 General Purpose probe P5934 Mictor probe P5960 DMAX probe Bandwidth Typical 2 Ghz 3dB 1 5 Ghz 3dB 2 Ghz 3dB Accuracy gain and offset Typical 80 mV 2 of signal amplitude Channel to channel...

Страница 17: ...ynchronously sampled and stored every 40 ps in a separate high resolution memory The storage speed can be changed by software to 80 ps 160 ps 320 ps or 640 ps with no loss in memory depth so that the...

Страница 18: ...gle channel Data valid window Setup and hold window 1 Typical P5910 General Purpose probe P5960 DMAX probe 260 ps single channel Mictor probe 500 ps single channel Data valid window Setup and hold win...

Страница 19: ...ck For data rates greater than 800 Mb s the TSM evaluates multiple data samples per TSM clock up to the maximum acquired data rate Counter timer range 2 fast no latency counter timers Maximum count is...

Страница 20: ...ing violation storage in either glitch external or Ts Th internal does not impact record length Time stamp bits and resolution duration 51 bits 20 ps 3 25 days Table 12 Symbolic support Characteristic...

Страница 21: ...t 29 7 cm 11 7 in Width 43 7 cm 17 2 in Overall dimensions Typical Depth 38 7 cm 15 2 in Weight Typical Includes instrument with front cover and empty accessory pouch TLA6401 13 52 kg 29 lbs 13 oz TLA...

Страница 22: ...Specifications 12 TLA6400 Series Product Specifications Performance Verification...

Страница 23: ...published specifications documented in the Specifications chapter of this manual The performance verification procedures certify the accuracy of an instrument and provide a traceability path to natio...

Страница 24: ...fixture Threshold Accuracy One required Refer to Threshold Accuracy Test Fixture See page 25 Test fixture Setup and Hold Minimum of two test fixtures required Refer to Setup and Hold Test Fixture See...

Страница 25: ...gnostics Prerequisites Warm up time 30 minutes Perform the following tests to complete the functional verification procedure 1 If you have not already done so power on the instrument and start the log...

Страница 26: ...rements specified Table 16 Parameters checked by verification procedures Parameter Verification method System clock CLK10 accuracy 1 Verified by the 10 MHz system clock test Threshold accuracy 1 Verif...

Страница 27: ...window and select System Configuration from the System menu 4 In the System Configuration dialog box select 10 MHz Clock from the list of routable signals in the External Signal Out selection box and...

Страница 28: ...the trigger parameters 1 Open the Setup window a In the Group column enter a name for the probe group that you are testing Test in the example b Define the signals for the group that you are testing c...

Страница 29: ...Performance verification procedures Figure 1 Setting trigger parameters TLA6400 Series Product Specifications Performance Verification 19...

Страница 30: ...Start the logic analyzer and verify that it does not trigger 7 Decrease the voltage in 10 mV steps waiting at least 3 seconds between steps to make sure that the logic analyzer continues to run withou...

Страница 31: ...ata pattern representing half the period of CH1 for example 001100110011 binary starting with 00 5 Connect the setup and hold test fixtures to the DTG channels that you have set up Connect 50 SMA term...

Страница 32: ...a Acquisition Length 1K or greater b Storage Options Normal 5 Create a new group right click in the Group Name column 6 Select Add Group from the pop up window Rename the new group Test 7 In the Prob...

Страница 33: ...ation procedures 8 Open the LA Trigger window and select the Power Trigger tab Set up three states as shown See Figure 2 Figure 2 Set the trigger states TLA6400 Series Product Specifications Performan...

Страница 34: ...f the DTG causes a false trigger on the logic analyzer each time you change the delay Then observe if the data is correct in the waveform window and ignore any false triggers Continue increasing the c...

Страница 35: ...ription Example part number Square pin strip 0 100 x 0 100 2 x 8 contacts or two 1 x 8 contacts SAMTEC part number TSW 102 06 G S Wire 20 gauge Soldering iron and solder 50 W Build procedure Use the f...

Страница 36: ...in strip 0 100 x 0 100 2 x 2 contacts or two 1 x 2 contacts SAMTEC part number TSW 102 06 G S SMA termination 50 2 GHz bandwidth Johnson part number 142 0801 866 SMA adapter Male to male Johnson part...

Страница 37: ...nnectors together See Figure 5 Figure 5 Solder the SMA connectors together 6 Solder the ground conductors of the SMA connectors together 7 Attach the termination and coupler to the fixture Figure 6 Co...

Страница 38: ...mber Certificate number Verification performed by Verification date Test data Characteristic Specification Tolerance Incoming data Outgoing data Clock frequency 10 MHz 500 Hz 9 9995 MHz 10 0005 MHz 2...

Отзывы: