![Tektronix TLA6400 Series Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/tektronix/tla6400-series/tla6400-series_technical-reference_1078215017.webp)
Speci
fi
cations
Table 8: Probing speci
fi
cations (cont.)
Characteristic
Description
Channel-to-channel skew
(Maximum)
±40 ps instrument only
Channel-to-channel skew (maximum)
P5910 general purpose probe
Instrument channel-to-channel skew plus ±125 ps
P5960 D-Max probe
Instrument channel-to-channel skew plus ±125 ps
P5934 Mictor probe
Instrument channel-to-channel skew plus ±500 ps
Channel-to-channel skew
(Typical)
±40 ps instrument only
P5910 general purpose probe
Instrument channel-to-channel skew plus ±40 ps
P5960 D-Max probe
Instrument channel-to-channel skew plus ±40 ps
P5934 Mictor probe
Instrument channel-to-channel skew plus ±250 ps
Hot-pluggable probes
Probes can be added or removed from the instrument while the instrument is turned on.
Minimum probe input voltage
300 mV
p-p
1
Characteristics are veri
fi
ed only at the 10 x attenuation
Table 9: Sampling speci
fi
cations
Characteristic
Description
MagniVu timing
Resolution
Data is synchronously sampled and stored every 40 ps in a separate high resolution memory.
The storage speed can be changed by software to 80 ps, 160 ps, 320 ps, or 640 ps (with no loss
in memory depth) so that the high resolution memory covers more time at a lower resolution.
Record length
128 k per channel
P5910 General Purpose
probe
P5934 Mictor probe
P5960 DMAX probe
Minimum pulse width
(Typical)
250 ps
500 ps
250 ps
Sampling (clocking) modes
Internal, External, Custom
Internal clocking
1x mode
1.25 ns (800 MS/s) maximum clocking period on all channels
2x mode
625 ps (1.6 GS/s) maximum clocking period on all channels
4x mode
312.5 ps (3.2 GS/s) maximum clocking period on half channels (using 2:1 demultiplex mode)
50 ms is the slowest clocking rate. A 1–2–5 sequence is supported, but it starts with 312.5 ps,
625 ps, 1.25 ns, 2.5 ns (PowerFlex), 5 ns, 10 ns, 20 ns, 50 ns, etc. It is possible to use storage
control and only store data when it has changed (transitional storage).
P5910 General Purpose
probe
P5934 Mictor probe
P5960 DMAX probe
Minimum pulse width
(Typical)
250 ps
500 ps
250 ps
TLA6400 Series Product Speci
fi
cations & Performance Veri
fi
cation
7
Содержание TLA6400 Series
Страница 8: ...Service safety summary vi TLA6400 Series Product Specifications Performance Verification...
Страница 10: ...Preface viii TLA6400 Series Product Specifications Performance Verification...
Страница 22: ...Specifications 12 TLA6400 Series Product Specifications Performance Verification...