Performance Veri
fi
cation
DC Attenuation Accuracy
This test checks the probe gain by measuring known voltages with a multimeter. The probe is then used to measure the
same voltages, and then a comparison calculation is made.
Preparation
1.
Set the vertical offset to 0 V.
2.
Connect the MicroCKT test tips to the DC source. Observe proper polarity: red to (+), black to (–).
3.
Connect a second DMM to the DC source.
Figure 9: Setup for differential-mode tests