Calibration
Test procedure
Check the Calibration Status of the instrument:
1.
From the Utilities menu, select Instrument
Calibration.
2.
In the Status box, check that
pass
is
displayed.
3.
If the Instrument Calibration Status is not
pass, run the Signal Path Compensation
routine: Disconnect all probes and signal
sources from the oscilloscope, and select
Calibrate. When
pass
displays in the
status box, proceed to the next step.
12
TDP0500 & TDP1000 Quick Start User Manual