Calibration
Calibration
The probe calibration routine minimizes your measurement errors by optimizing the gain and offset of the probe and oscilloscope
combination. We recommend that you repeat the probe calibration on each channel that you use. Individual calibration constants
are stored for each probe on each channel.
NOTE.
Probe calibration functions are not available on all oscilloscope models.
Prerequisites
The equipment must be warmed up for 20 minutes, and the calibration status of the host instrument must be pass.
Required equipment
The required equipment for calibration is the same as for the functional check. (See page 9,
TDP0500 & TDP1000 Quick Start User Manual
11