FFT
4.
Turn the
Vertical Scale
(volts/division) knob to ensure that the entire
waveform remains on the screen. The oscilloscope may display erroneous
FFT results (by adding high frequency components) if the entire waveform is
not visible.
5.
Turn the
Horizontal Scale
(seconds/division) knob to provide the resolution
you want in the FFT spectrum.
6.
If possible, set the oscilloscope to display many signal cycles.
If you turn the
Horizontal Scale
knob to select a faster setting (fewer cycles),
the FFT spectrum shows a larger frequency range, and reduces the possibility
of FFT aliasing. (See page 55,
.) However, the oscilloscope also
displays less frequency resolution.
To set up the FFT display, follow these steps:
1.
Push the
FFT
front-panel button to see the FFT side menu.
2.
Push
Source
from the side menu.
3.
Turn the
Multipurpose
knob to highlight the source channel. Push the knob
to select the channel.
NOTE.
Trigger and position any transient or burst waveforms as closely as
possible to the center of the screen.
Nyquist Frequency
The highest frequency that any real-time digitizing oscilloscope can measure
without errors is one-half the sample rate. This frequency is called the Nyquist
frequency. Frequency information above the Nyquist frequency is undersampled,
which causes FFT aliasing. (See page 55,
The math function transforms the center 2048 points of the time-domain
waveform into an FFT spectrum. The resulting FFT spectrum contains 1024
points that go from DC (0 Hz) to the Nyquist frequency.
Normally, the display compresses the FFT spectrum horizontally into 250 points,
but you can use the FFT Zoom function to expand the FFT spectrum to more
clearly see the frequency components at each of the 1024 data points in the FFT
spectrum.
NOTE.
The oscilloscope vertical response rolls off slowly above its bandwidth
(50 MHz, 70 MHz, 100 MHz, 150 MHz or 200 MHz, depending on the model, or
20 MHz when the Bandwidth Limit option is ON). Therefore, the FFT spectrum
can show valid frequency information higher than the oscilloscope bandwidth.
However, the magnitude information near or above the bandwidth will not be
accurate.
w w w . e k t