Performance veri
fi
cation
NOTE.
The following measurements are taken
with the sampling oscilloscope. The TekVPI
oscilloscope only provides power and offset to
the probe.
18.
On the sampling oscilloscope, expand
the horizontal scale to locate the step
edge: Set the vertical scale to 5 mV/div,
and adjust horizontal range to 100 ps/div
while maintaining the edge view. For a
more stable measurement display, turn
averaging on.
NOTE.
Do not touch the HF Probe Tip adapter
when making calibration measurements.
Measurement accuracy is degraded when the
probe tip adapter is handled.
TAP2500, TAP3500, and TAP4000 Instruction Manual
57