Basic operation
Probe offset
The probe offset is adjustable for operation within the linear range of the probe, and to increase the sensitivity of the probe at
higher DC measurement voltages. Using the offset to cancel DC signal components enables optimal probe performance. (See
Figure 1 on page 14.)
NOTE.
See your oscilloscope manual for speci
fi
c instructions on using the offset control.
To set the probe offset, follow these steps:
1.
Use the vertical position control to set a
zero reference level on the oscilloscope
display.
2.
Set the oscilloscope coupling to
DC
and
5 V/div. This sets the oscilloscope to
display the full offset dynamic range of the
probe.
TAP2500, TAP3500, and TAP4000 Instruction Manual
15
Содержание TAP2500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv TAP2500 TAP3500 and TAP4000 Instruction Manual...
Страница 32: ...Basic operation Figure 1 Dynamic and Offset Limitations 14 TAP2500 TAP3500 and TAP4000 Instruction Manual...
Страница 57: ...Specifications Figure 2 TAP2500 Typical bandwidth TAP2500 TAP3500 and TAP4000 Instruction Manual 39...
Страница 58: ...Specifications Figure 3 TAP3500 Typical bandwidth 40 TAP2500 TAP3500 and TAP4000 Instruction Manual...
Страница 59: ...Specifications Figure 4 TAP4000 Typical bandwidth TAP2500 TAP3500 and TAP4000 Instruction Manual 41...
Страница 80: ...Maintenance 62 TAP2500 TAP3500 and TAP4000 Instruction Manual...
Страница 82: ...Index T TekVPI 2 Test record 59 64 TAP2500 TAP3500 and TAP4000 Instruction Manual...