![Tektronix TAP2500 Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/tektronix/tap2500/tap2500_instruction-manual_1078181073.webp)
Performance veri
fi
cation
10.
Turn off the display for Channel 1 so that
only Channel 8 is shown on screen.
11.
Adjust the oscilloscope horizontal and
vertical position controls to display a signal
similar to that shown.
12.
Set the oscilloscope horizontal scale to
50 ps/div and center the waveform.
NOTE.
Do not touch the HF Probe Tip adapter
when making calibration measurements.
Measurement accuracy is degraded when the
probe tip adapter is handled.
13.
Use the oscilloscope measurement
capability to display rise time. Increase the
stability of the pulse edge measurement
by using averaging, if available. Rise time
is determined from the 10% and 90%
amplitude points on the waveform. Record
the rise time as
t
s
.
TAP2500, TAP3500, and TAP4000 Instruction Manual
55
Содержание TAP2500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv TAP2500 TAP3500 and TAP4000 Instruction Manual...
Страница 32: ...Basic operation Figure 1 Dynamic and Offset Limitations 14 TAP2500 TAP3500 and TAP4000 Instruction Manual...
Страница 57: ...Specifications Figure 2 TAP2500 Typical bandwidth TAP2500 TAP3500 and TAP4000 Instruction Manual 39...
Страница 58: ...Specifications Figure 3 TAP3500 Typical bandwidth 40 TAP2500 TAP3500 and TAP4000 Instruction Manual...
Страница 59: ...Specifications Figure 4 TAP4000 Typical bandwidth TAP2500 TAP3500 and TAP4000 Instruction Manual 41...
Страница 80: ...Maintenance 62 TAP2500 TAP3500 and TAP4000 Instruction Manual...
Страница 82: ...Index T TekVPI 2 Test record 59 64 TAP2500 TAP3500 and TAP4000 Instruction Manual...