![Tektronix RSA3303B Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/tektronix/rsa3303b/rsa3303b_user-manual_1078145030.webp)
Calibration
Figure 14: DC offset
To run the DC offset calibration, do the following:
1.
Press the
CAL
key on the front panel.
2.
Press the
Calibrate DC Offset
side key.
The calibration runs. It takes several seconds to complete the process.
Calibrating IF Flatness (RSA3408B Only)
The IF
fl
atness calibration adjusts the IF (intermediate frequency)
fl
atness using
the internal signal generator. It optimizes the
fl
atness of gain and phase within
the IF bandwidth automatically. This calibration is recommended in digital
modulation analysis.
To run the IF
fl
atness calibration, do the following:
1.
Press the
CAL
key on the front panel.
2.
Press the
Calibrate IF Flatness
side key.
The calibration runs. It takes several seconds to complete the process.
Display Brightness Adjustment
Adjust the brightness of the display according to your environment.