Installation
Figure 10: Spectrogram display
Powering Off the Analyzer
Turn off the power switch on the front panel.
CAUTION.
When you power on or off the analyzer, you must use the front-panel
power switch. Failure to do so may cause the operating system to shut down
improperly.
When you power on the analyzer again, wait at least 10 seconds after the last
power off.
When you press the front-panel ON/STANDBY switch, the analyzer starts a
shutdown process (including a Windows shutdown) to preserve settings and then
power off. The LED next to the power switch changes to orange. Avoid using the
rear-panel power switch or disconnecting the power cord to power off the analyzer.
To completely remove power to the analyzer, perform the shutdown just described,
and then set the power switch on the rear panel to off.