![Tektronix RSA3303B Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/tektronix/rsa3303b/rsa3303b_user-manual_1078145027.webp)
Calibration
Calibration
Perform these routines to optimize the analyzer performance:
Gain calibration
Center offset calibration
DC offset calibration
IF
fl
atness calibration
Display brightness adjustment
Each item is explained in this section.
Cal Menu
Use the
CAL
key to calibrate the analyzer.
Figure 11: Calibration menu structure
The Cal menu contains the following controls:
Calibrate All.
Performs all possible calibration operations.
Calibrate Gain.
Calibrates the internal gain steps. (See page 16,
Gain
Calibrate Center Offset.
This calibration cancels a center offset. (See page 17,
Calibrate DC Offset.
This calibration cancels DC offset in baseband. (See
page 17,
Calibrate IF Flatness.
Calibrates the IF (intermediate frequency)
fl
atness. (See
page 18,
Calibrating IF Flatness (RSA3408B Only)