Section 4—PS 503A
MAINTENANCE
AND
INTERFACING
INFORMATION
PREVENTIVE MAINTENANCE
There
are no special preventive maintenance
procedures that apply to
the PS 503A. Refer
to the power
module instruction
manual
for general
preventive
maintenance procedures and
instructions.
CORRECTIVE
MAINTENANCE
Refer
to the power module instruction manual for
general
corrective maintenance procedures and instruc
tions.
TROUBLESHOOTING
Use
the Performance Check, Adjustment Procedure,
and Circuit
Description
as aids to locate trouble in the
event
of
equipment failure. The test equipment listed in
the
Performance Check
and
Adjustment Procedures will
prove useful
in troubleshooting the PS 503A.
FUNCTIONS
AVAILABLE
AT
REAR
CONNECTOR
Pins are
available at the rear connector for routing
signals
to
and from the PS 503A for specialized
applications
(see Fig. 4-1, Input-Output Assignments for
Plug-In
Rear Interface Connector). One or more com
partments of a
multi-plug-in power module can be wired
with
barriers
installed to provide specific functions
between compartments.
See
power
module instruction
manual
for
additional
information.
FUNCTION
PROGRAMMING
OF
REAR
INTERFACE
CONNECTOR
Remote
Resistance Program
Remove
the
jumpers from
F - F (+Volts supply) and H -
H (—Volts supply).
Install jumpers between
E - E (+Volts
supply)
and G - G (—Volts supply).
Connect
a 10.0 kΩ
, 1% resistor between pins 28A and
28B (+Volts supply)
and pins 27A and 27B
on the interface
connector.
Install the PS
503A
into aTM 500 Series Power
Module.
Connect
a digital voltmeter between the
+20-volt
and common
output terminals and between the -20-volt
and
common
output
terminals. Adjust R45, +Adj and
R145,
-Adj to obtain a 20 V reading for each supply orthe
supply being
programmed. Turn
off the PS 503A and
remove
the 10.0 kΩ,
1% resistor.
NOTE
Do
not
turn on the PS 503A without a program
resistor
connected between
pins 28A and 28B
(+Volts
supply) and pins 27A
and
27B (—Volts
supply)
on the
interface connector or the over
voltage
protection circuit will cause
the fuse to blow.
The PS 503A has now been programmed at 500 G/volt
up
to
20 V, i.e„ a 1 kΩ change in the program resistor
results
in
a 2 V change in the PS 503A output.
A capacitor connected across the points
marked
C
f
+
or
C
f
–
may be needed
to stop oscillations caused by the
lead length associated with the program
resistor.
Remote
Voltage
Program
Dual Tracking.
Remove the
jumper from A - A, then
connect
a
jumper between B - B.
Connect
the
+ lead of the remote voltage to pin 20A and
the
- lead to pin 24A and 24B on the interface connector.
Install
the PS 503A into a TM 500 Series Power Module.
Connect
a
digital voltmeter between the
+20 V and
common output
terminals and between the —20 V and
common output
terminals.
Apply 9 V from the remote
voltage
source to the PS 503A.
Adjust R45, +Adj and
R145,
—Adj to obtain a 20 V reading for each supply.
4-1
Содержание PS 503A
Страница 1: ...Tektronix COMMITTED TO EXCELLENCE PS 503A DUAL POWER SUPPLY INSTRUCTION MANUAL...
Страница 4: ...Fig 1 1 PS 503A Dual Power Supply plug in module PS 503A...
Страница 36: ...Options PS 503A OPTIONS No options are available at this time...
Страница 41: ...I8S4 OX R V P FEB IR77 PS 5O3A...
Страница 42: ...is 4 qx FL V P FEB IR77 Poweb Supply J bm bowo t above...
Страница 45: ...TO...
Страница 46: ...1834 04...
Страница 47: ...PARTIAL POWER SUPPLY...
Страница 52: ...REV A DEC 1975 21...
Страница 53: ......