Adjustments
46
P6249 4 GHz 5X Active Probe Instruction Manual
1103
Power supply
--
+
Power supply
+
Digital multimeter
--
BNC-to-dual
banana adapter
50
Ω
Precision
terminator
Figure 16: P6249 offset range setup
4.
Set the DC power supply (using the DMM) to +2.5 VDC.
5.
Observe the DC value on the DMM.
6.
Adjust
Offset Range
for a DMM reading of 0 V
±
5 mV. Refer to
Figure 14 on page 43 for the location of the offset range
adjustment.
NOTE
. After the offset zero and offset range have been adjusted,
perform a probe calibration and functional check to verify perfor-
mance.