Performance Verification
38
P6249 4 GHz 5X Active Probe Instruction Manual
Rise Time
This procedure verifies that the probe meets rise time specifications.
The probe rise time is calculated from rise times obtained by
measuring the test system without the probe attached, and the test
system plus the probe.
1.
Connect the test equipment as shown in Figure 12.
2.
Set the oscilloscope trigger to internal clock.
3.
Turn on channel 1 on the SD-24, and then set the oscilloscope
vertical scale to 50 mV/div.
4.
Set the oscilloscope horizontal scale to 100 ps/div.
5.
Adjust the oscilloscope horizontal and vertical position controls
to display a signal similar to that shown in Figure 12.
11801
1103
Power supply
Calibration
step generator
BNC to SMA adapter
Output
Trigger
Input
Generator
remote head
Internal
clock output
Input
Precision coaxial cable
Figure 12: Test system rise time setup
6.
Use the oscilloscope measurement capability to display rise time.
Rise time is determined from the 10% and 90% amplitude points
on the waveform. Record the rise time as
t
s.
The system rise time (
t
s
) that you measured in step 6 represents the
rise time of the test system without the probe.
Содержание P6249
Страница 1: ...Instruction Manual P6249 4 GHz 5X Active Probe 071 0538 05 www tektronix com...
Страница 4: ......
Страница 8: ...Table of Contents iv P6249 4 GHz 5X Active Probe Instruction Manual...
Страница 28: ...Functional Check 16 P6249 4 GHz 5X Active Probe Instruction Manual...
Страница 34: ...Helpful Hints 22 P6249 4 GHz 5X Active Probe Instruction Manual...
Страница 40: ......
Страница 44: ...Theory of Operation 32 P6249 4 GHz 5X Active Probe Instruction Manual...
Страница 54: ...Performance Verification 42 P6249 4 GHz 5X Active Probe Instruction Manual...
Страница 64: ...Maintenance 52 P6249 4 GHz 5X Active Probe Instruction Manual...
Страница 68: ...Troubleshooting 56 P6249 4 GHz 5X Active Probe Instruction Manual...