Type
P6042
SECTION
2
OPERATING
INSTRUCTIONS
Change
information, if any, affecting this section is found at the rear of the manual.
General
The
P6042 Current Probe is designed to work into either a
50-ohm
system or a high
impedance oscilloscope using a
50-ohm termination.
This
section of the manual describes the
use
of
the
P6042 Current Probe. Information in this section
is
given
to
aid in making accurate measurements
with the
probe.
Handling
The
P6042 is designed to be as rugged as possible,
consistent with
good
high-frequency response and size. How
ever, as
with all
precision
devices, the probe and cable
should
be handled
with care
to avoid damage. Special care
should
be
taken that
the cable not be crushed, kinked, or
pulled.
Avoid
dropping the probe, as some of the most sen
sitive
circuitry is located within the
probe assembly.
Operating
Voltage
The P6042 can be powered from either a 115-volt or a 230-
volt nominal line
voltage source. The line voltage selector
assembly on
the
rear panel converts the instrument from
one
operating
range to the other. In addition, this switch
assembly
changes
the primary connections of
the power
transformer
to allow
selection
of one of three regulating
ranges.
The assembly also includes the two line fuses. When
the instrument is converted
from 115-volt to 230-volt nominal
operation, or
vice versa,
the switch assembly connects or
disconnects
one
of the fuses to provide the correct protec
tion
for the instrument. Use the following procedure to con
vert
this instrument between nominal
line voltages or reg
ulating ranges.
1.
Disconnect
the instrument from the power source.
2.
Loosen
the two captive
screws which hold the cover
onto
the
switch assembly; then pull to remove the cover.
3.
To convert
from
115-volt to 230-volt operation, pull out
the Voltage
Selector switch bar (see Fig. 2-1); turn it around
and plug into
the remaining holes. Change the power
cord
to match the
powersource receptacle
or use
a 115-volt to
230-volt plug adapter.
4.
To change regulating
ranges, pull out the Range Selec
tor
switch
bar (see Fig. 2-1); slide it to the
desired position
and
plug it back in. Select a range which is centered about
the
average line
voltage to which the instrument will be con
nected (see
Table
2-1).
5.
Re-install
the
cover and tighten the two captive screws.
6.
Before
applying power to the instrument, check that the
indicating
tabs on the switch bars are protruding through the
correct holes for
the
desired nominal line voltage and regulat
ing
range.
Fig.
2-1.
Line
Voltage
Selector assembly
on the rear
panel (shown
with
cover
removed).
CAUTION
The
P6042 should not be operated with the Volt
age
Selector and Range Selector
switches in the
wrong
position for the line voltage applied. Opera
tion
of
the instrument with the switches in the
wrong
positions will either provide
incorrect opera
tion
or
damage the instrument.
TABLE
2-1
Regulating
Ranges
Range
Selector
Switch
Position
Regulating
Range
115-Volts
Nominal
230-Volts
Nominal
LO (switch bar
in left
holes)
90
to
110
volts 180 to 220 volts
M (switch
bar in mid
dle
holes)
104
to
126
volts 208 to
252 volts
HI
(switch
bar in right
holes)
112
to 136
volts 224 to 272 volts
Operating
Temperature
The
P6042 depends upon
convection cooling to maintain
a
safe
operating temperature.
Adequate clearance on all
sides
must be
provided to allow heat to be dissipated away
from
the instrument.
Do not block or restrict the air flow
2-1
Содержание P6042 PROBE
Страница 3: ......
Страница 4: ...Characteristics Type P6042 Fig 1 1 P6042 Current Probe and accessories...
Страница 7: ...Characteristics Type P6042 1 3...
Страница 8: ...Characteristics Type P6042 1 4...
Страница 10: ...Operating Instructions Type P6042 Fig 2 2 Function of front panel controls and connectors 2 2...
Страница 28: ...NOTES...
Страница 30: ...Calibration Type P6042 Fig 6 1 Test equipment recommended 6 2...
Страница 46: ...NOTES...
Страница 64: ...HE RMAL COUPLER SEE PARTS LIST FOR SEMICONDUCTOR TYPES AMPLIFIER MRH S...
Страница 65: ...P 6 042 CURRENT PROBE DC TO 50MHz...
Страница 66: ...FIG 1 P6042 PROBE PACKAGE STANDARD ACCESSORIES P6042 PROBE...
Страница 67: ...FIG 2 P6042 PROBE MECHANICAL PARTS...
Страница 68: ...FIG 3 AMPLIFIER PO...
Страница 69: ...3 AMPLIFIER POWER SUPPLY P6042 PROBE...
Страница 70: ...FIG 4 CABINET...
Страница 71: ...FIG 4 CABINET P6042 PROBE...