Performance
Check—Type P6042
k. Change
CURRENT/DIV to 2 mA and check for
2.5
centimeters of
deflection.
7. Check Noise
a. Connect
the two GR
10X 50 Ω attenuators and the
current loop
to the square-wave generator Hi Amplitude
output.
Connect the probe to the current loop.
b.
Set
the CURRENT/DIV
switch to 1 mA, Time/cm to
.1 ms,
Triggering
Mode to Trig, and Stability clockwise to
display
two free-running traces.
c. Reduce
the
square-wave generator Amplitude
until the
two
traces just merge (the point at which the dark band
between
the
traces just disappears). See
Fig. 5-3B.
d. Switch
CURRENT/DIV to 10 mA and remove the two
10X
attenuators. Connect current loop and probe to the
square-wave
generator output.
e. Check the
amplitude of the display for
5 cm (
≤0.5
mA
+ 0.2 cm, tangentially measured).
See Fig.
5-3C.
f.
Remove
the probe from the square-wave generator
output
and place
in the holder (lock the slider).
8.
Check
Random
Trace Shift
a.
Set
CURRENT/DIV to
1 mA, degauss the
probe, and
center
the trace with the CURRENT/DIV BALANCE.
b.
Set the
controls
as follows:
Fig.
5-4.
Maximum
random
trace
shift
at
5
sec/cm,
stored.
Time-Base Plug-In
Horiz
Display
A
Mag
On
Time/cm
.2
µ
s
Trig
Mode
Auto
Slope
+
Coupling
AC
Source
Int
Storage
Oscilloscope
Upper
Screen Storage
Lower
Screen Storage
Norm
- Single Sweep
Triggering
Mode
Store
Store
Single
Sweep
Auto
Selectable
Bandwidth Plug-In
Low
Freq 3
dB Point
.1
High
Freq
3 dB Point
100
c. Push Erase
and
Reset button several times while adjust
ing
Intensity so that
the brightness of the trace is at mini
mum
storage
level.
d. Set Time/cm to
5 sec and push
Erase and Reset button.
e. Check stored display for
≤1.5 centimeters peak to peak.
See Fig.
5-4.
b.
Connect
the cable from the P6042 OUTPUT to the DC
to
100 MHz test oscilloscope Ch 1
Input, terminated in 50
ohms.
c.
Remove the
current loop from the square-wave genera
tor and
connect
to
the pulse generator output. Connect the
probe to
the
current loop. Connect
the delay cable to the
pulse
generator charge line connectors,
using the two lengths
of
RG-8A/U.
d.
Set
the pulse generator Voltage
Range
to 5.0, Pulse
Polarity to — and turn power on.
e. Adjust pulse
generator Amplitude and test oscilloscope
Trig
Level to obtain a 4-centimeter display. (Use the viewing
hood
if necessary).
f.
Check
aberrations for <3%,
peak to peak.
Remove the
probe
from
the pulse generator and
place in the holder.
Turn
X10
Mag off.
9. Check Aberrations
a.
Turn
the DC to 100 MHz test
oscilloscope on and set
the
controls as
follows:
Mode
Trigger
Volts/cm
Variable
Input
Coupling
Position
Vertical Plug-In
Ch
1
Norm
.2
Cal
DC
Trace
centered
10.
Check
AC Dynamic Range
a.
Connect
the
current loop to the constant amplitude
signal generator output.
b.
Set
the controls as follows:
DC
to 100 MHz Test Oscilloscope
Volts/cm
Time/cm
Input
Coupling
Vertical
Position
.5
20
µs
Gnd
trace
centered
5-4
Содержание P6042 PROBE
Страница 3: ......
Страница 4: ...Characteristics Type P6042 Fig 1 1 P6042 Current Probe and accessories...
Страница 7: ...Characteristics Type P6042 1 3...
Страница 8: ...Characteristics Type P6042 1 4...
Страница 10: ...Operating Instructions Type P6042 Fig 2 2 Function of front panel controls and connectors 2 2...
Страница 28: ...NOTES...
Страница 30: ...Calibration Type P6042 Fig 6 1 Test equipment recommended 6 2...
Страница 46: ...NOTES...
Страница 64: ...HE RMAL COUPLER SEE PARTS LIST FOR SEMICONDUCTOR TYPES AMPLIFIER MRH S...
Страница 65: ...P 6 042 CURRENT PROBE DC TO 50MHz...
Страница 66: ...FIG 1 P6042 PROBE PACKAGE STANDARD ACCESSORIES P6042 PROBE...
Страница 67: ...FIG 2 P6042 PROBE MECHANICAL PARTS...
Страница 68: ...FIG 3 AMPLIFIER PO...
Страница 69: ...3 AMPLIFIER POWER SUPPLY P6042 PROBE...
Страница 70: ...FIG 4 CABINET...
Страница 71: ...FIG 4 CABINET P6042 PROBE...