CAUTION.
To avoid damaging the input circuitry of the P5210 probe,
do not apply voltage that is more than 2,200 V
RMS
between either
input and earth ground or more than 4,400 V
RMS
between the inputs.
In addition, the peak voltage must be less than 5,600 V
(DC + peak AC)
between the inputs.
Above 1.5 MHz, the voltage limit decreases as frequency increases.
See Figure 2. The input limit applies to both the 100X and 1,000X
settings.
Voltage
(RMS)
1000V
100V
DC or 0
10M
1M
Frequency (Hz)
5M
50M
500V
50V
100k
500k
Category II maximum
voltage limit (2,200 V)
RF burn risk area
(shaded)
Voltage derating
with frequency
4000V
Figure 2: Safety Limits (voltage between either input and earth ground
)