Theory of Operation
Processor System
The Processor system contains a microprocessor that controls the entire
instrument. The processor system also contains FLASH ROM, system RAM, and
interfaces to USB ports and the Ethernet port.
Power Converter
The Mezzanine board converts the +12 V power to other voltages used for the
analog and digital circuitry throughout the system. The +5 V standby power is
used to keep the Front Panel powered up at all times that AC power is connected
to the unit.
Front-Panel Board
The Front Panel board contains a microprocessor that reads the front-panel
buttons and controls, sends this information to the processor system on the Main
board, and controls power to the system. The Front Panel board also generates
the probe compensation output signal, provides an interface to the application
modules, and controls power to the system.