Model TRX-1100V-CONN 3-Slot Male HV Triaxial Cable Connector Instructions
2
071368600 / September 2019
This triaxial connector is shipped with purple heat
-
shrink tubing. As shown in the previous figure, it should be
installed on the SC-22 cable. The purple marking indicates that the cable and connector assembly is rated for
up to 1500 V and conforms to IEC 61010-1 and IEC 61010-2-30 safety standards. The purple marking also
indicates that they can be attached to other connectors with purple markings, such as the FORCE and
SENSE connectors on the Model 2470 rear panel, shown in the following figure.
Figure 3: Model 2470 rear
-
panel purple connections
Do not attach male triaxial connectors without purple markings to female connectors with
purple markings. Triaxial connectors without purple markings are not rated for high voltage.
Mating incompatible connectors can cause damage to the connectors.
Dimensions
Figure 4: TRX-1100V-CONN connector dimensions