background image

Model TRX-1100V-CONN 3-Slot Male HV Triaxial Cable Connector Instructions 

071368600 / September 2019 

This triaxial connector is shipped with purple heat

-

shrink tubing. As shown in the previous figure, it should be 

installed on the SC-22 cable. The purple marking indicates that the cable and connector assembly is rated for 

up to 1500 V and conforms to IEC 61010-1 and IEC 61010-2-30 safety standards. The purple marking also 

indicates that they can be attached to other connectors with purple markings, such as the FORCE and 

SENSE connectors on the Model 2470 rear panel, shown in the following figure. 

Figure 3: Model 2470 rear

-

panel purple connections 

Do not attach male triaxial connectors without purple markings to female connectors with 

purple markings. Triaxial connectors without purple markings are not rated for high voltage. 

Mating incompatible connectors can cause damage to the connectors. 

Dimensions 

Figure 4: TRX-1100V-CONN connector dimensions 

Содержание Keithley TRX-1100V-CONN

Страница 1: ...SourceMeter Instrument It can be used for instrument testing that requires high voltage and high throughput combined with DC measurement capability and broad testing flexibility Figure 1 TRX 1100V CO...

Страница 2: ...0 2 30 safety standards The purple marking also indicates that they can be attached to other connectors with purple markings such as the FORCE and SENSE connectors on the Model 2470 rear panel shown i...

Страница 3: ...rmediate contact to outer body 2500 VRMS Operating environment 0 C to 50 C up to 70 relative humidity at 35 C SC 22 triaxial cable construction Figure 5 SC 22 triaxial cable dimensions SC 22 electrica...

Страница 4: ...mensions to ensure the correct insertion depth in the connector Excessive exposed conductor length presents electrical shock hazards to equipment and personnel If the strip length is too short imprope...

Страница 5: ...To prepare the SC 22 cable for assembly to the TRX 1100V CONN 1 Use the cable ringing tool or wire stripper to strip the cable to the proper dimensions Figure 7 Strip SC 22 cable 2 Remove the graphite...

Страница 6: ...gure 9 Add hex nut and V gasket to triax cable 2 Install the V clamp as shown in the following figure Carefully seat the V clamp into the V gasket when inserting The sharp edge of the V clamp must spl...

Страница 7: ...lamp onto the V clamp Figure 12 Add front clamp 5 Install the gasket clamp insulator and inner clamp as shown in the following figure Figure 13 Add gasket clamp insulator and inner clamp 6 Flare the e...

Страница 8: ...ntact Crimp the center contact to the center conductor cable as shown in the following figure Figure 15 Add slotted clamp support insulator and center contact 8 Slide the outer contact over the center...

Страница 9: ...ns 071368600 September 2019 9 9 Insert the terminated cable assembly into the rear of the connector body as shown in the following figure 10 Tighten the cable nut with a wrench to 3 39 Newton meters 3...

Страница 10: ...ts will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exercise extreme caution when a...

Страница 11: ...r components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component note that selected parts should be purchased only through Keithl...

Отзывы: