[ CARE AND USE MANUAL ]

CONTENTS

I.  

INTRODUCTION

II. 

GETTING STARTED

 

a.  eCord Installation

 

b.  Column Installation

 

c.  Column Equilibration

 

d.  Initial Column 

 

  Efficiency Determination                               

 

e.  Column QR Code 

 

f. XSelect 

XP

 Columns 

 

  — VanGuard Protection

 

g.  Replacing the 

 

  VanGuard FIT Cartridge

II. 

COLUMN USE

 

a.  Sample Preparation

 

b.  pH Range

 

c. Solvents

 

d. Pressure 

 

e. Temperature

III. 

COLUMN CLEANING, 

 

REGENERATING AND STORAGE

 

a.  Cleaning and Regeneration

 

b. Storage

IV. 

eCORD INTELLIGENT CHIP TECHNOLOGY

 

a. Introduction

 

b. Installation

 

c.  Manufacturing Information

 

d.  Column Use Information

V. 

ADDITIONAL INFORMATION

 

a.  Tips for Maximizing ACQUITY 

 

  HSS Column Lifetimes 

 

b.  Recommended Flow Rates and 

 

  Backpressures for Reversed-Phase 

 

  ACQUITY HSS Columns

VI. 

CAUTIONARY NOTE

I. INTRODUCTION

Thank you for choosing a Waters™ XSelect 

XP

 and/or XSelect 

Premier HSS Column. The XSelect HSS packing materials 
are manufactured in a cGMP, ISO 9001:2000 certified plant 
using ultra pure reagents. Each batch of XSelect HSS material 
is tested chromatographically with acidic, basic and neutral 
analytes and the results are held to narrow specification ranges 
to assure excellent, reproducible performance. Every column 
is individually tested and a Performance Chromatogram 
and Certificate of Batch Analysis are provided on the 
eCord™ intelligent chip.

The XSelect Premier HSS Column is offered with or without a 
VanGuard™ Fully-Integrated Technology [FIT] Cartridge. 
To address the desire to extend the operating lifetimes of 
analytical columns, the VanGuard FIT Cartridge is designed 
to prevent the non-desired introduction of sample matrix or 
particulates onto the column without degrading the separation. 
It can be easily replaced to restore separation performance and 
extend the analytical column’s lifetime. 

In addition, the XSelect Premier HSS Columns utilize 
MaxPeak™ High Performance Surfaces, an innovative 
technology designed to increase analyte recovery, sensitivity, 
and reproducibility by minimizing analyte/surface interactions 
that can lead to sample losses.

XSelect 

XP

 and XSelect Premier 2.5 µm HSS Columns

Содержание XSelect XP

Страница 1: ...ra pure reagents Each batch of XSelect HSS material is tested chromatographically with acidic basic and neutral analytes and the results are held to narrow specification ranges to assure excellent rep...

Страница 2: ...ents entry of air into the detection system 4 Gradually increase the flow rate as described in Step 2 5 Once a steady backpressure and baseline have been achieved proceed to the next section Note If m...

Страница 3: ...d to the inlet of the column For more information on VanGuard products visit www waters com vanguard g Replacing the VanGuard FIT Cartridge The VanGuard FIT Cartridge is designed specifically for XSel...

Страница 4: ...nd mobile phases are miscible in order to avoid sample and or buffer precipitation Filter sample with 0 2 m membranes to remove particulates If the sample is dissolved in a solvent that contains an or...

Страница 5: ...d cleaning call your local Waters office for additional support Table 3 Recommended pH and Temperature Limits for XSelect HSS Columns Column Particle Size Pore Diameter Surface Area pH Limits Temperat...

Страница 6: ...out of the bed Note If a column has been run with a mobile phase that contains formate e g ammonium formate formic acid etc and is then flushed with 100 acetonitrile slightly longer equilibration tim...

Страница 7: ...ncludes mobile phases running conditions and analytes used to test the columns In addition the QC results and acceptance is placed onto the column Figure 5 The eCord chip provides the user with an ove...

Страница 8: ...nal HPLC column and therefore mobile phases last much longer To reduce the chances of mobile phase contamination or degradation only prepare what you need for analysis or store excess bulk quantities...

Отзывы: