Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-18
2470-901-01 Rev. A /
May
2019
If you send this command without the
<function>
parameter, it will set autoclear for all measure
functions.
Example
:CALC2:VOLT:LIM1:CLE:AUTO ON
:CALC2:VOLT:LIM1:AUD FAIL
:CALC2:VOLT:LIM1:LOW 0.25
:CALC2:VOLT:LIM1:UPP 2.5
:CALC2:VOLT:LIMIT1:STAT ON
:READ?
:CALC2:VOLT:LIMIT1:FAIL?
Set limit autoclear on.
Enable the beeper for limit 1 when a voltage measurement
exceeds the limit.
Set lower limit 1 for voltage to 0.25 V.
Set upper limit 1 for voltage to 2.5 V.
Enable limit 1 testing for voltage.
Make a reading; the limit is checked and results display on the
front panel
Return the test results; example output if the test fails on the
low limit:
LOW
The test results are automatically cleared.
Also see
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
(on page 12-18)
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
This command clears the results of the limit test defined by
Y
.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<Y>
Limit number:
1
or
2
Details
Use this command to clear the test results of limit
Y
when the limit auto clear option is turned off. Both
the high and low test results are cleared.
To avoid the need to manually clear the test results for a limit, turn the auto clear option on.
If you send this command without the
<function>
parameter, it clears the limit for all measure
functions.