Series 2600B System SourceMeter® instrument User's Manual
Section 3: Instrument description
2600BS-900-01 Rev. A / August 2021
3-11
7. Ground
2601B, 2602B, 2604B, 2611B, 2612B, 2614B
Ground terminal for connection output HI or LO to
chassis ground.
Ground screw for connections to chassis ground.
2635B
Triaxial connector on ground module.
Screw terminal connector on ground module.
2634B, 2636B
Triaxial connector on ground module.
Screw terminal connector on ground module.
8. TSP-Link
Expansion interface that allows a 2600B and other
TSP-enabled instruments to trigger and communicate
with each other. Use a category 5e or higher LAN
crossover cable.
The TSP-Link is not available on the 2604B, 2614B,
and 2634B.
9. Power module
Contains the ac line receptacle and power line fuse. The
instrument can operate on line voltages of 100 V ac to
240 V ac at line frequencies of 50 Hz or 60 Hz.
10. RS-232
Female DB-9 connector. For RS-232 operation, use a
straight-through (not null modem) DB-9 shielded cable to
connect to the computer.