Models 2450-, 2460-, & 2461-EC Potentiostats And Galvanostats User's Manual Section 4: Potential Pulse & Square Wave
077110402/November 2016
4-7
To adjust the Pulse Width:
1. From the Pulse Settings screen, select
Pulse
Width
.
2. Enter a value. The Pulse Width can be no more than 99 percent of the Period setting.
Figure 71: Entering the Pulse Width
3. Select
OK
.
The Sample Time is the length of time over which the actual measurement is taken. The Sample
Time cannot exceed the Pulse Width minus 1 ms (PW - 1 ms).
The measurement is taken at the end of the pulse. A brief latency period between the end of the
measurement and the end of the pulse allows the analog-to-digital converter in your potentiostat to
complete the measurement.
To adjust the Sample Time:
1. From the Pulse Settings screen, select
Sample Time
.
2. Enter a value.
Figure 72: Entering a Sample Time
3. Select
OK
.