User service
5.
Set the TIVM series to the following:
Range:
2X
CLAMPING:
Off
Push the SELF CAL button to perform the self calibration (wait for the
SELF CAL Status indicator to turn solid green).
6.
On Channel 1 of the oscilloscope set the Vertical Scale to 200 mv/div.
7.
Connect the IVTIP1X sensor tip cable to the MMCX Jack (female) to BNC
female adapter.
8.
Enable the output of the Tek-DPG.
9.
The Channel 1 waveform should appear on the oscilloscope display and be
approximately vertically centered on the display. If not, ONLY adjust the
vertical settings as necessary to center the waveform on the display.
Measure the propagation
delay
Complete the following steps to measure and record the propagation delay.
1.
Set up the delay measurement on the oscilloscope as follows:
a.
Select
Measure
.
b.
Select
Add Measurement
.
c.
Select Measurement Type:
Delay
.
d.
Select Con
fi
gure:
Delay
.
e.
Set the Source to
R1
.
f.
Set Delay To to
CH1
.
g.
Select
OK Add Measurement
.
TIVM Series Measurement System User Manual
49
Содержание IsoVu TIVM Series
Страница 1: ...xx TIVM Series IsoVu Measurement System ZZZ User Manual P071349500 071 3495 00...
Страница 2: ......
Страница 6: ......
Страница 18: ...Preface xii TIVM Series Measurement System User Manual...
Страница 40: ...Operating information 22 TIVM Series Measurement System User Manual...
Страница 46: ...Application examples 28 TIVM Series Measurement System User Manual...
Страница 72: ...User service 54 TIVM Series Measurement System User Manual...
Страница 80: ...Appendix B Compliance information 62 TIVM Series Measurement System User Manual...