User service
Position:
0
(centered)
Offset:
-500 mV
Deskew:
0
seconds
5.
Set the Horizontal menu to the following setups:
Horizontal Scale to
10 ns/div
.
Horizontal Position:
40 ns
Acquire:
Average 128
6.
Set the Tek-DPG to the following setups:
0 to -1 1kHz mode
(Mode 1)
Output Enable:
ON
7.
The Channel 1 waveform should appear approximately on the center of the
oscilloscope display. If the waveform is not vertically centered in the display,
adjust the display as necessary to vertically center the waveform.
8.
Use the following steps to save the Channel 1 waveform as the reference
waveform (R1).
Press
MENU
on the oscilloscope.
Select
Save Waveform
.
Source:
CH1
.
Destination:
R1
.
Select
OK Save
to save the Channel 1 waveform as reference waveform
R1. The new reference waveform should now be displayed on the
oscilloscope.
9.
Disable the output of the Tek-DPG.
10.
Disconnect the Tek-DPG Deskew Pulse Generator output BNC cable from
the Channel 1 input of the TekVPI oscilloscope.
Create the TIVM series
waveform
Complete the following steps to set up the TIVM series waveform:
1.
Connect the Comp box of the TIVM series measurement system to Channel 1
of the TekVPI oscilloscope.
2.
Connect a IVTIP1X sensor tip cable to the TIVM series input (tighten the
SMA connector from the cable to the sensor head and attach the nose cone.)
3.
Allow the measurement system to warm up for 20 minutes at an ambient
temperature of 20 ° to 30 °C.
4.
Connect an MMCX Jack (female) to BNC female adapter to the output BNC
cable of the Tek-DPG.
48
TIVM Series Measurement System User Manual
Содержание IsoVu TIVM Series
Страница 1: ...xx TIVM Series IsoVu Measurement System ZZZ User Manual P071349500 071 3495 00...
Страница 2: ......
Страница 6: ......
Страница 18: ...Preface xii TIVM Series Measurement System User Manual...
Страница 40: ...Operating information 22 TIVM Series Measurement System User Manual...
Страница 46: ...Application examples 28 TIVM Series Measurement System User Manual...
Страница 72: ...User service 54 TIVM Series Measurement System User Manual...
Страница 80: ...Appendix B Compliance information 62 TIVM Series Measurement System User Manual...