![Tektronix IsoVu TIVM Series Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/tektronix/isovu-tivm-series/isovu-tivm-series_user-manual_1078002050.webp)
Reference information
Table 9: Electrical speci
fi
cations (cont.)
Characteristic
Description
Common mode rejection ratio (
Typical
)
Sensor tip cable/adapter
DC
100 MHz
200 MHz
500 MHz
1 GHz
IVTIP1X, 1X Sensor tip
cable
> 120 dB
120 dB
110 dB
100 dB
90 dB
IVTIP5X, 5X Sensor tip
cable
> 120 dB
120 dB
110 dB
100 dB
90 dB
IVTIP10X, 10X Sensor
tip cable
> 120 dB
120 dB
110 dB
100 dB
90 dB
IVTIP25X, 25X Sensor
tip cable
> 120 dB
110 dB
100 dB
100 dB
90 dB
IVTIP50X, 50X Sensor
tip cable
> 120 dB
100 dB
90 dB
90 dB
80 dB
MMCX-to 0.1 in
(2.54 mm) square pin
adapter with sensor tip
cable.
> 120 dB
70 dB
60 dB
40 dB
30 dB
MMCX-to 0.062 in
(1.57 mm) square pin
adapter with sensor tip
cable.
> 120 dB
70 dB
60 dB
40 dB
30 dB
Common mode voltage
range
2 kVpk (2 kV CAT I, 1000 V CAT II)
Common mode resistance
(
Typical
)
N.A. due to Galvanically isolation (
fi
ber optic connection)
Common mode
capacitance
3
(
Typical
)
~2 pF
Sensor tip cable/adapter
Overload indicator On
Sensor head input SMA
V
in
< -3 V or V
in
> +3 V
IVTIP1X, 1X Sensor tip cable
V
in
< -3 V or V
in
> +3 V
IVTIP5X, 5X Sensor tip cable
V
in
< -12 V or V
in
> +12 V
IVTIP10X, 10X Sensor tip cable
V
in
< -16 V or V
in
> +16 V
IVTIP25X, 25X Sensor tip cable
V
in
< -25 V or V
in
> +25 V
Overload indicator range
(
Typical
)
IVTIP50X, 50X Sensor tip cable
V
in
< -35 V or V
in
> +35 V
1
The difference between the measured DC gain and the nominal DC gain, divided by the nominal DC gain and expressed as a percent.
2
The input referred offset error voltage when the input is shorted and the probe input offset is set to 0 volts
3
The capacitance between the sensor head and a reference plane. The sensor head is placed six inches (15.25 cm) above the reference plane.
32
TIVM Series Measurement System User Manual
Содержание IsoVu TIVM Series
Страница 1: ...xx TIVM Series IsoVu Measurement System ZZZ User Manual P071349500 071 3495 00...
Страница 2: ......
Страница 6: ......
Страница 18: ...Preface xii TIVM Series Measurement System User Manual...
Страница 40: ...Operating information 22 TIVM Series Measurement System User Manual...
Страница 46: ...Application examples 28 TIVM Series Measurement System User Manual...
Страница 72: ...User service 54 TIVM Series Measurement System User Manual...
Страница 80: ...Appendix B Compliance information 62 TIVM Series Measurement System User Manual...